Remote photoacoustic measurements in aqueous solutions using an optical fiber

Publication Type

Journal Article

Date Published

12/1990

Authors

DOI

Abstract

A photoacoustic spectrometer was developed for remote optical absorption measurements in aqueous solutions using an 85‐m optical fiber to deliver pulsed tunable dye laser radiation to a sample cuvette located in a glove box. The spectrometer was tested using aqueous solutions of praseodymium ions. Beer’s law was verified down to a concentration of 8×10−6M for an equivalent absorptance of 3.2×10−5 cm−1.

Journal

Review of Scientific Instrumentation

Volume

61

Year of Publication

1990

Issue

12

Organization